Edwards Lifesciences

EV1000 Flo Trac Sensor Setup Guide

Setup Guide

2 Pages

FloTrac Sensor EV1000 Clinical Platform Setup Guide 6. Pressurize the Pressure Bag until it reaches 300 mmHg. 7. Fast-flush the FloTrac sensor and tap on tubing and stopcocks to remove any residual bubbles. 8. Connect tubing to arterial catheter, then aspirate and flush system to assure no residual bubbles remain. 9. Level the FloTrac sensor to the phlebostatic axis. Note: It is important to keep the FloTrac sensor level to the phlebostatic axis at all times to ensure accuracy of cardiac output. For professional use. CAUTION: Federal (United States) law restricts this device to sale by or on the order of a physician. See instructions for use for full prescribing information, including indications, contraindications, warnings, precautions and adverse events. Edwards Lifesciences devices placed on the European market meeting the essential requirements referred to in Article 3 of the Medical Device Directive 93/42/EEC bear the CE marking of conformity. Edwards, Edwards Lifesciences, the stylized E logo, EV1000, and FloTrac are trademarks of Edwards Lifesciences Corporation. © 2012 Edwards Lifesciences Corporation. All rights reserved. AR07457  Edwards Lifesciences I edwards.com USA I Switzerland I Japan I Singapore I Brazil  Note: FloTrac sensor setup is the same as a pressure transducer setup. If unfamiliar with pressure transducer setup, refer to step-by-step Sensor Setup. 1. Connect the power adapter and ethernet cable for both EV1000 panel and databox. Press the button on the panel. 2. When the boot up is complete, enter new patient data (patient ID, gender, age, height, and weight) or continue same patient. 3. If entering new patient data, use the touch screen to select and enter values. Press Home to continue. 4. Connect the FloTrac trifurcated databox cable to the back of the EV1000 databox. Then connect the green FloTrac connecting cable to the green capped connector on the FloTrac sensor. 5. Connect the bedside monitor’s arterial pressure cable to the red cable connector on the FloTrac sensor.
File Type: PDF
File Size: 386 KB
File Name: ev1000_flo_trac_sensor_setup_guide.pdf

By clicking Download you are confirming this is the correct document for your purposes
and that you agree to our Terms & Conditions