Endoplus

Coagulator Knot Pusher Probe Instructions for Use

2 Pages

Warning: Risk of injury: • The electrosurgical cord must cover the Monopolar Post. - (see Figure 1)  Figure 1 • DO NOT EXCEED maximum output voltage of 2000 Volts Peak at a Crest Factor < 1.6 • DO NOT EXCEED maximum output voltage of 2100 Volts Peak at a Crest Factor < 2.8 • DO NOT EXCEED maximum output voltage of 3000 Volts Peak at a Crest Factor < 5.5 • When applicable, the non-conductive Luer Port Cap must securely cover the luer port during use. • Neither the product nor the product container contains Natural Rubber Latex. Warning: Risk of fire: • DO NOT activate the electrode in the presence of flammable anesthetics or oxidizing gases (such as nitrous oxide (N2O) and oxygen) or in close proximity to volatile solvents (such as ether or alcohol), as explosion or fire may occur. • DO NOT place instrument near or in contact with flammable materials (such as gauze or surgical drapes). Instruments which are activated or hot from use may cause a fire. Warning: Risk of infection: • Devices ship from manufacturer non-sterile. Clean and sterilize per this IFU (Instructions for Use) prior to use. • Do not reuse devices which are suspected to have been contaminated with transmissible spongiform encephalopathy (TSE) agents (prions). Caution: • Federal (USA) law restricts this device to sale by or on the order of a physician. • Incorrect application of medical devices poses a risk of injury. Users of medical instruments must have an appropriate medical qualification and be acquainted with the application. • Laparoscopic surgery may result in gas embolism due to insufflation of gas into the abdomen. • Sharp edges or pinch points may damage personal protective equipment such as surgical gloves. • Due to the carcinogenic and infectious potential of electrosurgical by-products (such as tissue smoke plume and aerosols), protective eyewear, filtration masks, and effective smoke evacuation equipment should be used in both open and laparoscopic procedures. • Keep the active electrodes clean. Build-up of eschar may reduce the instrument’s effectiveness. Do not activate the instrument while cleaning. • Use instruments only in conditions that assure adequate visualization to minimize risk of misapplied electrosurgical energy. • Output power selected should be as low as possible for the intended purpose. • A singular instrument must be the sole conductor of energy to tissue. Do not conduct energy by touching an instrument to a second instrument contacting tissue. • Proper care and maintenance of hand-held laparoscopic instruments is essential for safe and effective operation. Prior to each use, instruments should be thoroughly examined for end of life indicators such as broken or worn parts which may inhibit the function. Specifically, instruments used for electrosurgery must be checked for nicks, cracks, or damaged insulation. Careful inspection upon receipt and frequent inspection during use for functional integrity is recommended as a safeguard against possible injury to patient or operator. Instruments should be used with extreme care when Inserting or removing from the cannula. Lateral pressure on the device during removal can damage the working tip, the shaft of the device, and/or insulation. • Where applicable, the non-conductive Luer Port Cap must be installed securely on the flush port during use to minimize insufflation leakage from the patient. The Luer Port Cap must NOT cover the flush port during cleaning and sterilization. • Use of instrument for a task other than that for which it is intended may result in a damaged or broken device. • Incorrect placement of trocars can adversely affect instrument effectiveness, patient safety, and operator comfort. • Electrosurgical Cords must comply with IEC 60601-2-2:2009. Always reference and follow the IFU (Instructions for Use) for Electrosurgical Cords. • Electrosurgical Generators must comply with IEC 60601-2-2:2009. Always reference and follow the IFU (Instructions for Use) for the ESU (Electrosurgical Generator).  Processing and Sterilization Instructions Pre-processing Instructions: • Initiate cleaning within 30 minutes of use and do not allow soil to dry prior to cleaning. • Inspect devices for end of life indicators. (See Product Life on p.3 for end of life indicators) Manual Cleaning Instructions for Suction Coagulators: 1. Prepare an enzymatic / neutral pH detergent in warm water (15°C - 30°C or 59°F to 86°F) per enzymatic / neutral pH detergent manufacturer’s IFU (Instructions for Use). 2. Flush devices with a minimum of 60mL of enzymatic/neutral pH detergent solution through the Luer Port. See Figure 2.  3. Completely immerse devices in the enzymatic/neutral pH detergent solution and allow devices to soak for a minimum of 5 minutes. 4. Ultrasonic clean device in enzymatic/neutral pH detergent solution for a minimum of 10 minutes. 5. Using a hand-held soft bristle brush, use a back and forth motion to brush all surfaces of devices with particular attention paid to monopolar posts if present, crevices, grooves, fittings, and joints. • External Brush Characteristics: Soft nylon (non-metallic) bristled brush; 1mm diameter brush with soft nylon (non-metallic) bristles can be used to clean the tips of coagulators. 6. Insert a brush into the luer port and the interior lumen a minimum of three times for cleaning prior to sterilization. • Lumen Brush Characteristics: 5mm diameter brush with soft nylon (non –metallic) bristles. 7. Flush devices with a minimum of 60mL of enzymatic/neutral pH detergent solution through devices. Repeat the flush process a minimum of two (2) times or until all fluid exiting the lumen is clear of soil. 8. If visible soil is detected during the final lumen flush, repeat ultrasonic brushing and flushing steps until all fluid exiting the lumen is clear of soil. 9. Rinse the devices by completely immersing in warm water for a minimum of 30 seconds to remove any residual detergent or debris. Flush devices with a minimum of 60ml of warm tap water through the luer port. Repeat the flush process a minimum of two (2) times. 10. Manipulate devices to allow rinse water to drain from the lumen. 11. Visually examine each device for cleanliness and repeat cleaning procedure if visible soil remains.  Manual Cleaning Instructions for Non-Suction Coagulators and Knot Pushers: 1. Prepare an enzymatic / neutral pH detergent in warm water (15°C - 30°C or 59°F to 86°F) per enzymatic / neutral pH detergent manufacturer’s IFU (Instructions for Use). 2. Completely immerse devices in the enzymatic/neutral pH detergent solution and allow devices to soak for a minimum of 5 minutes. 3. Ultrasonic clean devices in enzymatic/neutral pH detergent solution for a minimum of 10 minutes. 4. Using a hand-held soft bristle brush, use a back and forth motion to brush all surfaces of devices with particular attention paid to monopolar posts, crevices, grooves, fitting, and joints. • External Brush Characteristics: 1mm diameter brush with soft nylon (non-metallic) bristles can be used to clean the tips of coagulators and knot pushers. 5. Rinse the devices by completely immersing in warm water for a minimum of 30 seconds to remove any residual detergent or debris. 6. Visually examine each device for cleanliness and repeat cleaning procedure if visible soil remains.  Inspection/Maintenance Instructions for All Devices: Proper care and handling is essential for satisfactory performance of any surgical device. The steps in this (IFU) Instructions for Use should be taken to ensure long and trouble-free service from all your surgical devices. Inspect devices before each use for end of life indicators. Specifically, instruments used for electrosurgery must be checked for nicks, cracks, gaps, or damage to shaft insulation. Careful inspection upon receipt and frequent inspection during use for end of life indicators and functional integrity is recommended as a safeguard against possible injury to patient or operator. If any conditions appear which would indicate that the device is not safe for use, do not use the device and return the device to Endoplus for investigation. Lubricate devices with moving parts prior to sterilizing by immersing for a minimum of 30 seconds in instrument milk or a steam permeable/water soluble lubricant. Allow devices to drip dry approximately 3 minutes following lubrication and prior to packaging for sterilization. Follow the lubricant manufacturer's IFU (Instructions for Use). Do not use silicon, mineral oil, or petroleum-based lubricants.  Sterilization Instructions: Flush Port Cap must be open during sterilization. Configure devices so that water pooling does not occur. Prevacuum Steam Sterilization Parameters if Sterilized in Instrument Wrap: Minimum Preconditioning Pulses: 3 Minimum Temperature: 132°C (270°F) Minimum Exposure Time: 4 minutes Minimum Dry Time: 20-30 minutes Sterilization Configuration: Wrapped (2 layer 1-ply or 1 layer 2-ply) Storage Instructions if Sterilized in Instrument Wrap: Devices must remain wrapped and be stored in a clean, dry environment to maintain sterility. Follow the Instrument Wrap manufacturer's IFU (Instructions for Use). Prevacuum Steam Sterilization Parameters if Sterilized in a Rigid Container: Minimum Preconditioning Pulses: 3 Minimum Temperature: 132°C (270°F) Minimum Exposure Time: 4 minutes Minimum Dry Time: 0 minutes Sterilization Configuration: Devices in a Laparoscopic Rack which is inside of a Rigid Container  Storage Instructions if Sterilized in a Rigid Container:  Figure 2  Suction Coagulator  2  Component Key 1. Luer Port 2. Monopolar Post  1  Devices must remain sealed in the rigid container, which has been validated to maintain sterility. Follow the Rigid Container manufacturer's IFU (Instructions for Use) for shelf life.  Product Life: Expected life is greatly dependent upon the care, handling, and use of the devices. Product life may be reduced by improper handling, excessive force, or third-party modifications. Use until devices exhibit end of life indicators. End of life indicators include: 1. Cracked, broken, bent, missing, loose, or damaged components. 2. Damaged, missing, or modified shaft insulation. 3. Intermittent electrical performance 4. Impaired mechanical function 5. Missing or Unreadable UDI  2 4  31  F-7.15.4 Rev A
File Type: PDF
File Size: 223 KB
File Name: Endoplus - Coagulator Knot Pusher Probe Instructions for Use - Rev A.pdf

By clicking Download you are confirming this is the correct document for your purposes
and that you agree to our Terms & Conditions