Intuitive Surgical
5 x 250 mm Curved Cannula, Endoscope Left, da Vinci Xi
da Vinci Xi Reprocessing Wall Chart
1 Page
![da Vinci Xi Reprocessing Xi ENDOSCOPE Inspect 7 Flush Flush all ports with pressurized cold water. 6 x 20 sec Connector Cover (if present) 2 Clean Wipe thoroughly using disinfectant, dry, and attach cover. 8 Spray Submerge and spray gap and base. 2 x 20 sec Solution 3 Prepare Prepare enzymatic solution (pH 7-11). 9 Brush Brush submerged or under running cold water. 60 sec 4 Prime Prime button ports, input discs, and base with enzymatic solution. Rinse 10 Final Rinse with critical water. 60 sec 5 Soak Soak endoscope and cable. 15 min 11 Inspect Inspect entire endoscope for soil. 6 Rinse Rinse under running cold water. 60 sec Chemical Disinfection (optional) 12 Apply per manufacter’s instructions and rinse. 1 Inspect entire endoscope for damage. da Vinci Universal Reprocessing INSTRUMENTS AND STAPLER 8 Rinse Rinse under running cold water. 60 sec Solution 2 Prepare Prepare enzymatic solution (pH 7-11). Cleaning 9 Ultrasonic Prime and clean in ultrasonic bath. 15 min 3 Prime Prime primary port with enzymatic solution. 10 Flush Flush ports with pressurized cold water. 20 sec 4 Soak Soak the device. 30 min Rinse 11 Final Rinse with critical water. 60 sec 20 sec 12 Inspect Inspect entire device for soil. 6 Spray Submerge and spray tip. 30 sec Thermal Disinfection (optional) 13 Disinfect per hospital policy. 7 Brush Brush submerged or under running cold water. 60 sec 1 Remove Accessories Remove all accessories before cleaning. 5 Flush Flush all ports with pressurized cold water. ACCESSORIES GROUP 1 Cannulae, Obturators, Release Kits, Hasson Cone, Instrument Introducers, Gage Pins, Suction Irrigator Kit, etc. Ultrasonic Cleaning (optional) 6 Clean in ultrasonic bath. 15 min Solution 2 Prepare Prepare enzymatic solution (pH 7-11). Rinse 7 Final Rinse with critical water. 10 sec 3 Soak Soak and actuate moving parts. 10 min 8 Inspect Inspect entire device for soil. 4 Rinse Rinse under running cold water. 20 sec Thermal Disinfection (optional) 9 Disinfect per hospital policy. 5 Brush Brush submerged or under running cold water. 60 sec 1 Remove Accessories Remove all accessories before cleaning. ACCESSORIES GROUP 2 Cautery Cord, Stapler Cable, PK Cords, Light Guide Cable, etc. 1 Disconnect Cords or Cables Rinse 5 Final Rinse with critical water. Disconnect all cords and cables from device. Solution 2 Prepare Prepare enzymatic solution (pH 7-11). 6 Inspect Inspect entire accessory for soil. 3 Wipe Wipe thoroughly using enzymatic solution. Thermal Disinfection (optional) 7 Disinfect per hospital policy. 4 Brush Brush submerged or under running cold water. 60 sec 60 sec Intuitive Surgical, Inc. 1266 Kifer Road Sunnyvale, CA 94086 USA Customer Service 800.876.1310 Copyright © 2017 Intuitive Surgical, Inc. All rights reserved. Intuitive®, Intuitive Surgical®, da Vinci® S™, da Vinci® Si™, da Vinci® X™, da Vinci® Xi™, and EndoWrist® are registered trademarks of Intuitive Surgical, Inc. Other brand and product names are trademarks or registered trademarks of their respective holders. PN 554059-01 Rev. A For Intuitive Surgical da Vinci® X/Xi™ Surgical System Instruments, Stapler, and Endoscopes. This chart is not intended to replace an in-service or a manual. Refer to da Vinci X/Xi Reprocessing Instructions and da Vinci X/Xi Stapler Reprocessing Instructions.](https://public.bioclinicalservices.com.au/library/representations/proxy/eyJfcmFpbHMiOnsiZGF0YSI6MTM4MTEzLCJwdXIiOiJibG9iX2lkIn19--6e40f6b2afd90be4070d0951837eb69e188bbc82/eyJfcmFpbHMiOnsiZGF0YSI6eyJyZXNpemUiOiIzMjB4In0sInB1ciI6InZhcmlhdGlvbiJ9fQ==--4214356078d1f110e449cda512a9432562da3e70/Intuitive%20Surgical%20-%20Intuituve%20Sugical%20-%20PN%20554059-01%20-%20da%20Vinci%20Xi%20Reprocessing%20Wall%20Chart%20-%20Rev.%20A.pdf.png)
Preview
Page 1
da Vinci Xi Reprocessing Xi ENDOSCOPE Inspect
7 Flush Flush all ports with pressurized cold water.
6 x 20 sec
Connector Cover (if present) 2 Clean Wipe thoroughly using disinfectant, dry, and attach cover.
8 Spray Submerge and spray gap and base.
2 x 20 sec
Solution 3 Prepare Prepare enzymatic solution (pH 7-11).
9 Brush Brush submerged or under running cold water.
60 sec
4 Prime Prime button ports, input discs, and base with enzymatic solution.
Rinse 10 Final Rinse with critical water.
60 sec
5 Soak Soak endoscope and cable.
15 min
11 Inspect Inspect entire endoscope for soil.
6 Rinse Rinse under running cold water.
60 sec
Chemical Disinfection (optional) 12 Apply per manufacter’s instructions and rinse.
1
Inspect entire endoscope for damage.
da Vinci Universal Reprocessing INSTRUMENTS AND STAPLER
8 Rinse Rinse under running cold water.
60 sec
Solution 2 Prepare Prepare enzymatic solution (pH 7-11).
Cleaning 9 Ultrasonic Prime and clean in ultrasonic bath.
15 min
3 Prime Prime primary port with enzymatic solution.
10 Flush Flush ports with pressurized cold water.
20 sec
4 Soak Soak the device.
30 min
Rinse 11 Final Rinse with critical water.
60 sec
20 sec
12 Inspect Inspect entire device for soil.
6 Spray Submerge and spray tip.
30 sec
Thermal Disinfection (optional) 13 Disinfect per hospital policy.
7 Brush Brush submerged or under running cold water.
60 sec
1
Remove Accessories Remove all accessories before cleaning.
5 Flush Flush all ports with pressurized cold water.
ACCESSORIES GROUP 1 Cannulae, Obturators, Release Kits, Hasson Cone, Instrument Introducers, Gage Pins, Suction Irrigator Kit, etc.
Ultrasonic Cleaning (optional) 6 Clean in ultrasonic bath.
15 min
Solution 2 Prepare Prepare enzymatic solution (pH 7-11).
Rinse 7 Final Rinse with critical water.
10 sec
3 Soak Soak and actuate moving parts.
10 min
8 Inspect Inspect entire device for soil.
4 Rinse Rinse under running cold water.
20 sec
Thermal Disinfection (optional) 9 Disinfect per hospital policy.
5 Brush Brush submerged or under running cold water.
60 sec
1
Remove Accessories Remove all accessories before cleaning.
ACCESSORIES GROUP 2 Cautery Cord, Stapler Cable, PK Cords, Light Guide Cable, etc.
1
Disconnect Cords or Cables
Rinse 5 Final Rinse with critical water.
Disconnect all cords and cables from device.
Solution 2 Prepare Prepare enzymatic solution (pH 7-11).
6 Inspect Inspect entire accessory for soil.
3 Wipe Wipe thoroughly using enzymatic solution.
Thermal Disinfection (optional) 7 Disinfect per hospital policy.
4 Brush Brush submerged or under running cold water.
60 sec
60 sec
Intuitive Surgical, Inc. 1266 Kifer Road Sunnyvale, CA 94086 USA Customer Service 800.876.1310 Copyright © 2017 Intuitive Surgical, Inc. All rights reserved. Intuitive®, Intuitive Surgical®, da Vinci® S™, da Vinci® Si™, da Vinci® X™, da Vinci® Xi™, and EndoWrist® are registered trademarks of Intuitive Surgical, Inc. Other brand and product names are trademarks or registered trademarks of their respective holders. PN 554059-01 Rev. A
For Intuitive Surgical da Vinci® X/Xi™ Surgical System Instruments, Stapler, and Endoscopes. This chart is not intended to replace an in-service or a manual. Refer to da Vinci X/Xi Reprocessing Instructions and da Vinci X/Xi Stapler Reprocessing Instructions.