![Service Notes Healthcare No. 03 11/2011 CLARIO AC, CLARIO AC/DC – REPLACEMENT OF MEMBRANE Clario Airway Suction Pump](https://public.bioclinicalservices.com.au/library/representations/proxy/eyJfcmFpbHMiOnsiZGF0YSI6MTYyMjAyLCJwdXIiOiJibG9iX2lkIn19--af53bb32a133d27d6bf1b03112662b6f0775dfbc/eyJfcmFpbHMiOnsiZGF0YSI6eyJyZXNpemUiOiIzMjB4In0sInB1ciI6InZhcmlhdGlvbiJ9fQ==--4214356078d1f110e449cda512a9432562da3e70/Replacement%20Clario%20Membrane.pdf.png)
Preview
Page 1
Service Notes
Healthcare No. 03 11/2011 CLARIO AC, CLARIO AC/DC – REPLACEMENT OF MEMBRANE
Clario Airway Suction Pump
Introduction
In the past, we noticed that the requirement to replace defective Clario membranes was steadily increasing. However, so far this was not possible because spare parts and appropriate tools were not available. Until now, the only way to fix a defective membrane was to exchange the entire drive unit (Art. No. 200.0334 or 200.0335). With this service note we would like to instruct you how to handle and replace defective membranes.
Who is authorized to exchange the Clario Membrane? –– Service engineers who have the required equipment (removal pliers and mounting device). –– They need to have the relevant training, knowledge and practical experience to be able to guarantee a correct and professional repair.
Which Clario Membrane can be exchanged? On Clarios with S/N 1163227 or higher piston rod and m embrane are screwed together. Replacement of the m embrane is possible.
On Clarios with S/N 1163226 and lower, the piston rod and membrane are made from one piece. It is not possible to replace the single membrane. As before, use "Spare Part Housing with motor/Membrane"
Which Clario Membrane can be exchanged? In order to replace the membrane, you must use the following tools: left:
Clario locking ring removal pliers Item: 200.4926
right: Clario locking ring mounting device Item: 200.4928
Field Action
1. Disassembly (for further Instructions also refer to the Service Manual)
A
Remove the lock ring.
B
Use the pliers according to the photo to
Other tools are not recommended as they
Split the ring on two different positions.
might damage the housing or membrane.
It will be easier to remove the locking ring.
Remove the two screws from the bottom housing and remove the bottom housing.
use too much force! You might damage the housing.
split the locking ring.
C
A quick squeeze might help but do not
D
Unscrew the membrane from the piston rod. Dispose of the defective membrane.
Remove the PCB (Follow ESD protective instructions). Remove the four screws and remove the motor with the membrane.
2. Replacement of the Membrane
B
Reinsert the motor and membrane into
Tighten the motor with four screws.
D
Put the Clario on the assembly tool.
Position the lock ring.
F
Press the locking ring on the housing until
A
Take the new membrane and screw it to
C
E
the piston rod.
The marking line must be vertical.
the housing.
it clicks. Remove the housing from the assembling tool. Then assemble PCB and bottom housing.
3. Test function Check using optional vacuum gauge: A
Insert vacuum gauge in connection on the canister lid
B
-73kPa -550mmHg
-17kPa -125mmHg
Switch on device
-33kPa -250mmHg
C
Set vacuum to maximum -73kPa -550mmHg
-17kPa -125mmHg -33kPa -250mmHg
D
Compare vacuum values
+ 2000 m +1000 m + 500 m 0m
– 458 mmHg – 518 mmHg – 551 mmHg – 595 mmHg
– 61 kPa – 69 kPa – 74 kPa – 80 kPa
Tolerance - 15 %
4. Spare parts 099.0252 Spare Part Clario Membrane complete (including locking ring, membrane, screw) HC 2011_01/ROD
Medical Vacuum Technology for Healthcare Professionals
Local contact:
Please contact us or your local Medela representative for details.
Medela AG Lättichstrasse 4b 6341 Baar / Switzerland
Medela Australia Pty Ltd Medical Technology 3 Arco Lane Heatherton, Vic, 3202 Australia Phone +61 (0) 3 9552 8600 Fax +61 (0) 3 9552 8699 [email protected] www.medela.com.au
Medela Canada Inc. 4160 Sladeview Cres., Unit #8 Mississauga, Ontario L5L 0A1 Canada Toll Free 800 435 8316 Toll Free Fax 800 995 7867 [email protected] www.medela.ca
Medela UK Huntsman Drive Northbank Industrial Park Irlam, Manchester M44 5EG UK Phone +44 870 950 5994 Fax +44 870 389 2233 [email protected] www.medela.co.uk
Medela Inc. USA P.O. Box 660 1101 Corporate Drive McHenry, IL 60050 USA Phone +1 877 735 1626 Fax +1 815 363 2487 [email protected] www.medelasuction.com
©Medela AG/2011-11/A
www.medela.com