Medline Industries

Konig Surgical Instruments General Instrumentation Recommended Cleaning and Sterilization Instructions

2 Pages

General Instrumentation – Recommended Cleaning and Sterilization Instructions  Automated Washer: Neutral pH Detergent 1. 2.  Follow Pre-Cleaning steps outlined above. Load instruments into automatic washer per manufacturer’s recommended orientation. 3. Wash instruments per Mechanical Washer Parameters in Table 1. Table 1 Phase Time Description Detergent Pre-Wash 1 2 min Pre wash with cold tap None water Enzyme Wash 1 min Enzyme spray and soak Enzymatic with hot tap water Detergent Cold Tap 15 sec Cold tap water rinse (2x) None Water Rinse Wash 1 2 min Detergent wash with hot Neutral pH tap water Cleaner Rinse 1 Hot 15 sec Hot tap water rinse None Tap Water Pure Rinse 10 sec Hot purified water None Drying 7 min Hot air dry None  Pre Cleaning 1. 2. 3. 4. 5.  6.  All devices should be cleaned in the open position to allow solution to contact all surfaces. Contaminated instruments should be cleaned as soon as possible. Rinse off device to remove any excess gross soil. Submerge instruments in an enzymatic/ neutral pH detergent bath and allow soaking between 5 and 10 minutes. Use a soft bristled brush and gently remove any visible soil still remaining on the device. Be sure to clean hinges, crevices and other difficult to reach areas. Lumens should be cleaned with a soft bristled pipe cleaner of corresponding width and length to ensure the entire lumen has been scrubbed. Rinse instruments in purified water for a minimum of 2 minutes. Flush lumens, hinges, crevices and other difficult to reach areas until the water exiting the device is clear of soil and detergent. If soil still remains, repeat the steps above.  4.  Manual Cleaning 1. 2.  Rinse under cool running tap water to remove gross soil. Bathe in enzymatic detergent per manufacturer’s recommendation using lukewarm tap water for 1 minute. 3. Scrub thoroughly with a soft bristled brush to remove soil. Pass a stylet through lumens a minimum of 3 times and, using a syringe, aggressively flush lumens with enzymatic detergent to remove soil. 4. Rinse under cool running tap water and aggressively flush lumens with a syringe to remove detergent residuals. 5. Bathe in a neutral detergent per manufacturer’s recommendation using warm tap water for 3 minutes. 6. Scrub thoroughly with a soft bristled brush to remove soil. Pass a stylet through lumens a minimum of 3 times and, using a syringe, aggressively flush lumens with neutral detergent to remove soil. 7. Rinse under running reverse osmosis/deionized (RO/DI) water to remove detergent residuals. 8. Sonicate in enzymatic detergent per manufacturer’s recommendation for 10 minutes. 9. Rinse under running RO/DI water and aggressively flush lumens with a syringe. 10. Dry with a disposable, lint free cloth. 11. Visually inspect for cleanliness. Repeat cleaning process, as necessary, until visually clean.  Automated Detergent 1. 2. 3. 4.  5.  3. 4.  5. 6.  Washer:  Low  Repeat cleaning process, as  Impingement/High  pH  Follow Pre-Cleaning steps outlined above. Load instruments into automatic washer per manufacturer’s recommended orientation. Wash instruments per Mechanical Washer Parameters in Table 1. Low motor setting (low impingement) requires Neutral pH Cleaner to be replaced with Alkaline Detergent with pH above 11 in Wash 1 Phase. Visually inspect for cleanliness. Repeat cleaning process, as necessary, until visually clean.  Note: Ensure instruments are lubricated prior to sterilization and after last rinse cycle using a water soluble product intended for surgical instruments.  Sterilization Sterilization should be performed in a medical grade instrument tray or disposable paper or plastic pouch. Make certain that the instrument container is sealed in an appropriate packaging for sterilization. Sterilize in compliance with the local guidelines for hospital hygiene. Sterilization of instruments may be accomplished by Autoclave or Ethylene Oxide. Time and temperature parameters required for sterilization vary according to type of sterilizer, cycle design, and packaging material.  Ultrasonic Cleaning 1. 2.  Visually inspect for cleanliness. necessary, until visually clean.  Follow Pre-Cleaning steps outlined above. Submerge instruments fully opened in Ultrasonic Washer with cold distilled water and the minimum effective concentration of enzymatic cleaner per manufacturer’s recommendation. Ultrasonically clean instruments at 45kHz for 10 minutes. Rinse under cool running RO/DI water for 2 minutes and aggressively flush lumens with a syringe until water exiting instrument is clear of detergent. Dry with a disposable, lint free cloth. Visually inspect for cleanliness. Repeat cleaning process, as necessary, until visually clean.  Autoclave Sterilization • Testing should be conducted by each healthcare facility to ensure that the specific configuration of instrument sets is acceptable for the sterilization process. • Do not sterilize instruments at temperatures over 141°C (285°F). • All ring handled instruments must be autoclaved in the fully open position to prevent cracking of the box lock. • All instruments must be sterilized in the completely open and disassembled (i.e. taken-apart) configuration. Note that  2
File Type: PDF
File Size: 180 KB
File Name: Medline Industries - Konig - Konig Surgical Instruments General Instrumentation Recommended Cleaning and Sterilization Instructions - 2015.pdf

By clicking Download you are confirming this is the correct document for your purposes
and that you agree to our Terms & Conditions