ResMed

ResMed AirFit N30 Disinfection Guide Nov 2019

Disinfection Guide

3 Pages

Disinfection guide  Clinical use only  This guide is intended for multi-patient use of the AirFit™ N30 nasal mask in a sleep lab, clinic or hospital. If you use the mask as a single user in the home, refer to the User Guide for cleaning instructions. The processor is responsible for ensuring that reprocessing is completed in accordance with ResMed’s validated procedures.  Cleaning Mild Alkaline, Anionic Detergent e.g. Alconox  Neodisher Mediclean Forte  Thermal Disinfection Mask component1  1 2  Manual  Washer-Disinfector  Validated number of cycles  • Cushion  30  • Standard frame  30  • QuietAir™ frame2  –  • Split-band headgear  –  –  – 30  May not be available in all regions. For full details regarding the correct use of this mask, please refer to the User Guide. For a list of available replacement parts, check the Product Guide on ResMed.com Not intended for multi-patient re-use.  A  Split-band headgear  1  Buckle  2  Buttonhole  3  End cap  B  Cushion  C  QuietAir™ / Standard frame  4  Stabiliser  5  QuietAir vent  6  Vent  7  Mask tubing  English  1
File Type: PDF
File Size: 186 KB
File Name: ResMed AirFit N30 Disinfection Guide Nov 2019.pdf

By clicking Download you are confirming this is the correct document for your purposes